QCM Flow Cell Kit

The Resonant frequency of quartz changes when material attaches
the electrode's surface. This product is capable of super-micro
quantitative analysis by using this unique behavior.

solution: PBS with 2mM Dithiodipropionic acid
a: stationary state
Flow rate:
b:85,
c:110,
d:140,
e:160,
f:190 µl/min
Base line is stable enough as figure in right shows.
Above figure shows result of change in frequency with time by adding
:
avidin(2mg/ml), ethanol amine(5mM),
BSA in PBS(1%),
Biotin-Allo A lectin(0.05mg/ml, in 1%BSA),
galactose-carrying polymer(1mg/ml, in 1%BSA),
PBS
Quartz Crystal

Thickness of 10MHz quartz is 0.16 mm. To accelerate the oscillation frequency means to make the quartz thinner and delicate. Resolution and easiness to care for should be taken into the consideration. We can offer two surface conditions of quartz - Polished and Unpolished. Polished one has a 100 Å thickness of gold layer on a titanium adhensive layer. We would recommend unpolished quartz if there is possibility that the titanium layer might effect on your experiment. We would like you to note that the gold layer on unpolished quarts peals off easier than polished one because it does not have adhensive layer.
QCM Flow Cell Kit - Components
| Catalog No. | Description |
| 011343 | QCM Flow cell kit |
| Components | Quantity |
| Acrylic Flow Cell | 1 |
| PEEK Cell | 1 |
| PEEK Cap | 1 |
| Teflon Tubing | 1 meter |
| Dynaseal PEEK | 2 |
| Fixing Screw | 2 |
| Silicone O-Ring | 2 |
| Flow Cell Holder | 1 |
| Pt Counter Electrode | 1 |
This cell even can be used in two ways;
QCM Mode
Using physical absorption such as antigen-antibody reaction on a solid surface
![]() |
![]() |
EQCM Mode - Part code: 012028
Using electrochemical absorption
![]() |
![]() |