SEC2000 - Spectroelectrochemical flow cell
The system is composed for spectrometer, light source,
electrochemical analyzer, and spectroelectrochemical flow cell.
Working electrodes (ITO, platinum mini-grid, and gold mini-grid) are
replaceable. Reference electrode is an aqueous RE-3V reference
electrode (Ag/AgCl) or non-aqueous RE-7V reference electrode system
(Ag/Ag+). Working
electrode’s grid distance is 150 µm, and line width is 100 µm. Gold
electrode is very soft, when you clean the electrode surface with
Kim wiper, it could damage the electrode surface, and so you may
need to handle carefully.

SEC-F Spectroelectrochemical flow cell
| Catalog No. | Description |
| 011320 | SEC-F Spectroelectrochemical flow cell |
| Components | |
| 011330 | SEC-F flow cell block (A) |
| 011331 | SEC-F flow cell block (B) |
| 011329 | SEC-F flow cell base |
| 010562 | 74-UV UV/VIS Collimating Lens, 200-2000 nm |
| 011495 | SEC-FT500 Teflon gasket (2 pcs) |
| Dynaseal PEEK (3 pcs) | |
| Optional Products | |
| 011496 | SEC-F Pt grid electrode for flow cell |
| 011497 | SEC-F Au grid electrode for flow cell |
| 012179 | SEC-F Carbon grid electrode for flow cell |
| 011519 | SEC-F ITO electrode for flow cell (4 pcs) |
| 011810 | SEC-F 1000 µm probe type of optical fiber, 1 m |
| 012169 | RE-3V Reference electrode Screw type (Ag/AgCl) |
| 012173 | RE-7V Non Aqueous reference electrode (Ag/Ag+) Screw type |
| 011959 | SEC-FS500 Teflon gasket (2 pcs) |
Example of the simulation using SEC-F Spectroelectrochemical flow cell
In situ system response measurement using SEC-F cell was
performed. CV with simultaneous absorbance measurement was performed
in 10 mM Ferrocene in acetonitrile solution (Fig. 1). Formation of
the reaction product and intermediate product, during CV
measurement, was observed in real-time measurement of the absorbance
(Fig. 2). And, Fig. 3 shows the absorbance peak.
![]() |
![]() |
| Figure 1 Cyclic voltammogram of 10 mM Ferrocene in acetonitrile solution. |
Figure 2 Influence on the reagent and product during CV measurement. |
Cyclic voltammogram (0 to 0.8 V; scan rate 50 mV/s), with
simultaneous spectrum measurement, was performed using ITO electrode
in Ferrocene solution. Oxidation and Reduction peaks were observed
at 450 and 620 nm, respectively, as shown below (Fig.3).
![]() |
| Figure 3 Oxi-Red spectra of Ferrocene during CV measurement. |
SEC2000 - Combo kit
| Catalog No. | Description |
| 012196 | SEC2000-UV/VIS Spectrometer combo kit |
| Components | |
| 012191 | SEC2000-UV/VIS Spectrometer |
| 012193 | SEC2000-DH UV Light Source |
| 012195 | SEC2000-CUV Cuvette Holder |
| Catalog No. | Description |
| 012197 | SEC2000-VIS/NIR Spectrometer combo kit |
| Components | |
| 012192 | SEC2000-VIS/NIR Spectrometer |
| 012194 | SEC2000-TH Visible Light Source |
| 012244 | SEC2000-CUV Cuvette Holder with diffusion plate |



